Z-0695-2008

Smiths Medical, Add-On Kids Kit , Blood Sampling System

Abbott Medical / initiated 2007-11-21 / Terminated / root cause: packaging

Smiths Medical Asd Inc began a recall of Smiths Medical, Add-On Kids Kit , Blood Sampling System, Product Code: MX4037, Smiths Medical ASD, Inc, Dublin, Ohio 43016 on . The reason FDA records is "Misbranded; The product codes are labeled incorrectly with a green promotional label for "New split septum sampling site. Only use needle-free access devices .......", however this product code contains a sampling site...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0695-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Smiths Medical, Add-On Kids Kit , Blood Sampling System, Product Code: MX4037, Smiths Medical ASD, Inc, Dublin, Ohio 43016
Product code
CBT Arterial Blood Sampling Kit
Reason for recall
Misbranded; The product codes are labeled incorrectly with a green promotional label for "New split septum sampling site. Only use needle-free access devices .......", however this product code contains a sampling site which is accessed with a shrouded needle, NOT with a needle-free access device
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Smiths Medical mailed certified letters (Urgent Product Advisory Notifications) dated November 19, 2007, to all customers that received the affected lot. Firms are instructed to make the field correction (extra label removal) or mail the product back to Smiths for the correction.
Quantity in commerce
1440 units
Distribution
Nationwide including the states of CA, DC, IL, IN, MD, MN, NC, NE, SC, and TN and Puerto Rico.
Product codes and lots
Lot Numbers: 1189082, 1231444, 1178100, and 1225588.