Irix Gamma Camera, Scintillation (gamma) camera, Philips Medical System, Cleveland, OH 44143
Philips Medical Systems (Cleveland) Inc began a recall of Irix Gamma Camera, Scintillation (gamma) camera, Philips Medical System, Cleveland, OH 44143 on . The reason FDA records is "Resistor Failure: The Rotate Motion Shunt Resistor may overheat and results in the appearance of smoke, a burning smell, and an electronic stop condition, which will disable the gamma camera rotate motion and all other...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1511-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Irix Gamma Camera, Scintillation (gamma) camera, Philips Medical System, Cleveland, OH 44143
- Product code
- KPS System, Tomography, Computed, Emission
- Reason for recall
- Resistor Failure: The Rotate Motion Shunt Resistor may overheat and results in the appearance of smoke, a burning smell, and an electronic stop condition, which will disable the gamma camera rotate motion and all other motions.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Consignees were sent an Urgent Field Safety Notice on 1/30/08. The letter instructs users who experience a failure to remove the system from service and contact the recalling firm for immediate correction. The recalling firm will schedule appointments with users to replace the affected part beginning in March 2008. For additional information, contact 1-800-722-9377.
- Quantity in commerce
- Unknown at this time sinc the recall was expanded.
- Distribution
- Worldwide Distribution --- USA including states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, and WV, Puerto Rico and countries of Argentinia, Austrailia, Austria, Belgium, Brazil, Canada, China, Czech Republic, Denmark, Egypt, Finland, France, Germany, Iceland, India, Ireland, Israel, Italy, Japan, Netherlands, Norway, Philipines, Poland, Singapore, South Africa, Spain, Switzerland, and United Kingdom.
- Product codes and lots
- Model Numbers: 210857 and 210881. 4535 679 46991, or N210857 Irix System, 3/8" Crystal 4535 665 13551, or N210881 Irix System 3/4" Crystals 4535 679 46981, or N211038 Irix System 3/8 Precision 4535 665 13571, or N211040 Irix System 3/4 Precision (Note: Due to a part number scheme change at the firm, either number is representative of the identified system)