Siemens, Oncor Impression, Part No.: 58 57 920, Medical Charge particle radiation therapy system,with...
Siemens Medical Solutions USA, Inc began a recall of Siemens, Oncor Impression, Part No.: 58 57 920, Medical Charge particle radiation therapy system,with pre-R2.0 Optivue Flat Panel Positioners, Siemens Medical... on . The reason FDA records is "Product's collision detection system may not be sensitive enough to trigger a stop action without a significant amount of force". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0382-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens, Oncor Impression, Part No.: 58 57 920, Medical Charge particle radiation therapy system,with pre-R2.0 Optivue Flat Panel Positioners, Siemens Medical, Concord, CA 94520
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Product's collision detection system may not be sensitive enough to trigger a stop action without a significant amount of force.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm has issued a Customer Safety Advisory Notice dated 12/6/2006, to all sites that have the pre-R2.0 OPTIVUE Flat Panel Positioners installed. Users will have a new microswitch-based collision detection system, installed free of charge.
- Quantity in commerce
- 21 units
- Distribution
- Worldwide Distribution- USA including MA, ME, NJ, NY, PA, MD, WV, NC, SC, GA, FL, AL, TN, KY, OH, IN, MI, IA, WI, SD, ND, MT, IL, NE, LA, OK, TX, WY, NV, CA, AK AND AR, and countries of Serbia, Australia, Austria, Belgium, Canada, Denmark, El Salvador, France, Germany, Hungary, India, Ireland, Italy, Japan, Kuwait, Malaysia, Mexico, Netherlands, New Zealand, Norway, China, Philippines, Poland, Portugal, Qatar, Republic of Korea, Russia, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Turkey, Ukraine, UK.
- Product codes and lots
- Model number 5857920, Serial Numbers 3729 3345 3811 3845 3883 3973 3974 4021 4048 70-4083 70-4083 70-4143 70-4152 70-4165 70-4167 70-4207 70-4235 70-4257 70-4281 70-4296 70-4344