Z-0384-2008

Siemens, Mevatron M2 Primus, Part No.: 19 40 035, Mid-Energy Medical Charge particle radiation therapy...

Siemens Medical Solutions USA, Inc. / initiated 2007-03-23 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of Siemens, Mevatron M2 Primus, Part No.: 19 40 035, Mid-Energy Medical Charge particle radiation therapy system, with pre-R2.0 Optivue Flat Panel Positioners... on . The reason FDA records is "Product's collision detection system may not be sensitive enough to trigger a stop action without a significant amount of force". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0384-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens, Mevatron M2 Primus, Part No.: 19 40 035, Mid-Energy Medical Charge particle radiation therapy system, with pre-R2.0 Optivue Flat Panel Positioners, Siemens Medical, Concord, CA 94520
Product code
IYE Accelerator, Linear, Medical
Reason for recall
Product's collision detection system may not be sensitive enough to trigger a stop action without a significant amount of force.
Root cause tag
design
FDA root cause
Device Design
Action
The firm has issued a Customer Safety Advisory Notice dated 12/6/2006, to all sites that have the pre-R2.0 OPTIVUE Flat Panel Positioners installed. Users will have a new microswitch-based collision detection system, installed free of charge.
Quantity in commerce
9 units.
Distribution
Worldwide Distribution- USA including MA, ME, NJ, NY, PA, MD, WV, NC, SC, GA, FL, AL, TN, KY, OH, IN, MI, IA, WI, SD, ND, MT, IL, NE, LA, OK, TX, WY, NV, CA, AK AND AR, and countries of Serbia, Australia, Austria, Belgium, Canada, Denmark, El Salvador, France, Germany, Hungary, India, Ireland, Italy, Japan, Kuwait, Malaysia, Mexico, Netherlands, New Zealand, Norway, China, Philippines, Poland, Portugal, Qatar, Republic of Korea, Russia, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Turkey, Ukraine, UK.
Product codes and lots
Model number 1940035, Serial Numbers 3736 3756 3491 3849 3885 4087 70-4249 70-4286 70-4323