Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0713-2008

Coloplast Elefant Suction/Irrigation Cannula for Laparoscopy Part Numbers: ASP 105, ASP 150, ASP 145, ASP...

MEDLINE INDUSTRIES, LP - Northfield / initiated 2007-11-12 / Terminated / root cause: manufacturing process

Coloplast Corp began a recall of Coloplast Elefant Suction/Irrigation Cannula for Laparoscopy Part Numbers: ASP 105, ASP 150, ASP 145, ASP 180, Coloplast Corp., Minneapolis, MN 55411 on . The reason FDA records is "Lack of sterilization assurance: Sterilization of Elefant Suction-Irrigation Devices is not performed according to ISO 11135 standard". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0713-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Coloplast Elefant Suction/Irrigation Cannula for Laparoscopy Part Numbers: ASP 105, ASP 150, ASP 145, ASP 180, Coloplast Corp., Minneapolis, MN 55411
Product code
FQH Lavage, Jet
Reason for recall
Lack of sterilization assurance: Sterilization of Elefant Suction-Irrigation Devices is not performed according to ISO 11135 standard.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Coloplast sent customers an "Urgent: Recall Notification" letter on 11/12/07. The letter described that the sterilization of the products was not performed according to ISO 11135 standard. The notification requested the customers to return the product immediately and to send a fax or contact the company via email in regard to inventory at hand. A phone script for distributors also included an advice to discontinue dispensing the product and to fax or email in regard to inventory at hand.
Quantity in commerce
154
Distribution
Nationwide, including Puerto Rico.
Product codes and lots
All un-expired lots are being recalled.   ASP 105 ASP 145: Lot Serial # 7030294 ASP14502001 ASP14504101517  ASP14504023142 ASP1455022514 ASP1454110199 ASP14503019 ASP14502001 ASP14503019 ASP14503017 ASP14503019 ASP14503017 ASP14502001 ASP14503017 ASP14502001 ASP14504023142 ASP14504101517 ASP14504023142 ASP14504101517 ASP14504110200 ASP1455022514 ASP1454110199 ASP1455031783 ASP1454110199 ASP1455031783 ASP1455041217 ASP1454110199 ASP1455041217 ASP1455052183 ASP1455041217 ASP1455033506 ASP1455041217 ASP1455052183 ASP1455041217 ASP1454110199 ASP1455041217 ASP1455052183 ASP1454110199 ASP1455052183 ASP 150: Lot serial # ASP15002001 ASP 180: Lot Serial #7044270 7044267 04041932