Z-1118-2008

Fisher Healthcare Sure-Vue Serum/Urine hCG-STAT; in vitro diagnostic test; Catalog Number: FHC-A202; Product...

Innovacon, Inc. / initiated 2007-10-01 / Terminated / root cause: labeling and UDI

Innovacon Inc began a recall of Fisher Healthcare Sure-Vue Serum/Urine hCG-STAT; in vitro diagnostic test; Catalog Number: FHC-A202; Product Codes: FHC-A202-KFI3O & FHC-A202-KFI5O; Inverness... on . The reason FDA records is "Incorrect results: The Sure-Vue Serum/Urine hCG-STAT test, when interpreted at extended read times, may exhibit sensitivity to patient samples containing hCG at levels well below the cut-off, potentially resulting in a...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1118-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Innovacon, Inc.
Product
Fisher Healthcare Sure-Vue Serum/Urine hCG-STAT; in vitro diagnostic test; Catalog Number: FHC-A202; Product Codes: FHC-A202-KFI3O & FHC-A202-KFI5O; Inverness Medical Professional Diagnostics, San Diego, CA 92121.
Product code
JHI Visual, Pregnancy Hcg, Prescription Use
Reason for recall
Incorrect results: The Sure-Vue Serum/Urine hCG-STAT test, when interpreted at extended read times, may exhibit sensitivity to patient samples containing hCG at levels well below the cut-off, potentially resulting in a false positive interpretation by the user.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
Inverness Medical sent an Urgent Medical Device Alert letter, dated October 1, 2007, to all Innovacon customers for the product via US Postal Service return receipt requested. The field correction notice removed the extended read time of 10 minutes. Customers were instructed to note that although the package insert currently allows interpretation up to 10 minutes, do not read at times greater than 3 minutes for urine and 5 minutes for serum.
Quantity in commerce
1,825,480 units
Distribution
Nationwide
Product codes and lots
Lot Numbers: hCG6120203, hCG6120204, hCG7010080, hCG7010081, hCG7010038, hCG7010039, hCG7010040, hCG7030144, hCG7030145, hCG7030146, hCG7040076, hCG7040077, hCG7040078, hCG7040079, hCG7050121, hCG7050122, hCG7060073, hCG7060216, hCG7070042, hCG7070043, hCG7070044, hCG7007045, hCG7070060, hCG7070061, hCG7070186, hCG7070187, hCG7070191, hCG7080182, hCG7090049, hCG7090050, and hCG7090052.