Coloplast Elefant Suction/Irrigation Cannula for Laparoscopy with Tube, Part Numbers: ASP 165, ASP 170...

MEDLINE INDUSTRIES, LP - Northfield / initiated 2007-11-12 / Terminated / root cause: manufacturing process

Coloplast Corp began a recall of Coloplast Elefant Suction/Irrigation Cannula for Laparoscopy with Tube, Part Numbers: ASP 165, ASP 170, Coloplast Corp., Minneapolis, MN 55411 on . The reason FDA records is "Lack of sterilization assurance: Sterilization of Elefant Suction-Irrigation Devices is not performed according to ISO 11135 standard". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0714-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Coloplast Elefant Suction/Irrigation Cannula for Laparoscopy with Tube, Part Numbers: ASP 165, ASP 170, Coloplast Corp., Minneapolis, MN 55411
Product code
FQH Lavage, Jet
Reason for recall
Lack of sterilization assurance: Sterilization of Elefant Suction-Irrigation Devices is not performed according to ISO 11135 standard.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Coloplast sent customers an "Urgent: Recall Notification" letter on 11/12/07. The letter described that the sterilization of the products was not performed according to ISO 11135 standard. The notification requested the customers to return the product immediately and to send a fax or contact the company via email in regard to inventory at hand. A phone script for distributors also included an advice to discontinue dispensing the product and to fax or email in regard to inventory at hand.
Quantity in commerce
559
Distribution
Nationwide, including Puerto Rico.
Product codes and lots
All un-expired lots are being recalled.   ASP 165: Lot Serial# 04023897 6120192 7021217 7053185 7062896 7062897 7062903 7070016 ASP16502002 ASP16502005 ASP16503040 ASP16503043 ASP16503101459 ASP16503110460 ASP16504023143 ASP16504023897 ASP16504033803 ASP16504061589 ASP16504064478 ASP16504070805 ASP16504074192 ASP1655021255 ASP1655030739 ASP1655031634 ASP1655042461 ASP1655051481 ASP1655052172 ASP1655052174 ASP1655052175 ASP1655052651 ASP1655053628 ASP1655060297 ASP1655060299 ASP1655062276 ASP1655062277 ASP1655070732 ASP1656100365 ASP1656100366 ASP1656120191 ASP1657010843 ASP1657021217 ASP 170: Lot Serial # ASP1701000