Nexframe Stereotactic System Kits, Model Number(s): DB-1021-MR, DB-1031, DB-1040-BL, DB-1040-ST, DB-1041...
Medtronic Image Guided Neurologics, Inc. began a recall of Nexframe Stereotactic System Kits, Model Number(s): DB-1021-MR, DB-1031, DB-1040-BL, DB-1040-ST, DB-1041, DB-1041-BL, DB-1041-ST, DB-1042, DB-1042-BL... on . The reason FDA records is "Sterilty (package integrity) Compromised: Some failures were for damage to the outer pouch, while another set of failures were for the seals on this pouch. The seal between the inner tray and lid has not been...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0849-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Image Guided Neurologics, Inc.
- Product
- Nexframe Stereotactic System Kits, Model Number(s): DB-1021-MR, DB-1031, DB-1040-BL, DB-1040-ST, DB-1041, DB-1041-BL, DB-1041-ST, DB-1042, DB-1042-BL, DB-1042-ST, DB-1043, DB-1043-BL, DB-1043-ST, DB-2031, DB-2040-BL, DB-2040-ST, DB-2041, DB-2041-BL, DB-2041-ST, DB-2042, DB-2042-BL, DB-2042-ST, DB-2043, DB-2043-BL, DB-2043-ST,, Medtronic, Inc, Minneapolis, Minnesota
- Product code
- HAW Neurological Stereotaxic Instrument
- Reason for recall
- Sterilty (package integrity) Compromised: Some failures were for damage to the outer pouch, while another set of failures were for the seals on this pouch. The seal between the inner tray and lid has not been compromised and the contents remain sterile, however, the sterility of the outer surfaces of the inner tray and lid cannot be assured.
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- On 10/4/2007, consignees were notified by Medtronic field personnel, either by personal visits or telephone. All wholesalers of this product are OUS. Medtronic OUS Regulatory personnel communicated this issue directly to the wholesalers, who provided the letters to their customers. Medtronic field personnel are responsible for completing and returning reply cards confirming all notifications (100%).
- Quantity in commerce
- 385
- Distribution
- Worldwide, USA, Taiwan, Thailand, Japan, Spain, Sweden and The Netherlands.
- Product codes and lots
- Lot #s: 070380, 061248, 070175, 070375, 070621, 070825, 070634, 6510106A, 046510106A, 46515606, 046520506A, 046521606A, 046523406A, 61262, 70160, 070326, 070171, 60215, 070719, 070378, 070169, 070320, 70543, 070172, 610127, 610133, 70169, 61208, 70313, 51067, 60174, 60215 60578, 046510106C, 70153, 60175, 046510106F, 46534905, 046520506B, 070164, 070633, 070376, 070375, 070538, 070724, 070376, 61264, 61212, 070855, 070854, 070830, 070863, 070726, 070628, 60542, 60573, 60574, 60659, 61262, 70312, 070379, 61216, 70312, 070628, 070726, 61264, 070723 070166, 070629, 070164, 070379, 070856, 60583, 61237, 61238, 606110, 70312, 070159, 70174, 70144, 070160, 70375, 070631, 070825, 070726, 070628, 070863, 61208, 70145, 610134, 70321, 61247, 70315, 70334, 070724 070720, 070602, 070326, 070602, 070631, 070325, 070719, 70123, 70160, 046502607B, 070165, 070826, 046505707B, 60434, 60433, 046510106E