Z-0588-2008
Baxter Single Day INFUSOR Portable Elastomeric Infusion System
Baxter Healthcare Corporation / initiated 2007-11-29 / Terminated / root cause: manufacturing process
Baxter Healthcare Corp. began a recall of Baxter Single Day INFUSOR Portable Elastomeric Infusion System, Product Code: 2C1071KJP; 2mL/hr with Tubing Flow Restrictor; 12 units per case; Baxter... on . The reason FDA records is "Leaking: leaks at the tubing flow restrictor connection to either the coupler or the male luer during filling". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0588-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Baxter Single Day INFUSOR Portable Elastomeric Infusion System, Product Code: 2C1071KJP; 2mL/hr with Tubing Flow Restrictor; 12 units per case; Baxter Healthcare Corporation, Deerfield, IL 60015
- Product code
- MEB Pump, Infusion, Elastomeric
- Reason for recall
- Leaking: leaks at the tubing flow restrictor connection to either the coupler or the male luer during filling.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Baxter on 11/30/07 sent an Urgent Product Recall letter to affected customers, to the attention of the Director of Nursing with copies to the Materials Manager, via first class mail to advise them of the potential leaks during filling or use of the infusor. The accounts were instructed to immediately discontinue use and segregate the listed affected lot numbers, and contact Baxter Center for Service at 1-888-229-0001 to arrange for the return and credit for the affected product. Any clinical/technical questions were directed to the Baxter Healthcare Product Information Center at 1-800-933-0303.
- Quantity in commerce
- 86,192 units
- Distribution
- Worldwide: USA, Canada, Austria, Belgium, Denmark, Finland, France, Germany, Greece, Italy, Portugal, San Marino, Spain, Sweden, United Kingdom, South Africa, Colombia, Argentina, Hong Kong, Australia and New Zealand.
- Product codes and lots
- Lot Numbers: 07E037, 07E038, 07E039, 07E055, 07E056, 07E057, 07F015, 07F029, 07F030, 07F039, 07G003, 07G004, 07G043, 07H004, 07H005, 07H006, 07H035, 07H036, 07H037, 07H057, 07H059, 07H061, 07H074, 07J016, 07J018, 07J052, 07J054, and 07K020