Z-1065-2008
Non-continuous ventilator accessories under the following brand names: a) REMStar (Tower) Ultrafine Qty 1...
Respironics, Inc. / initiated 2007-10-15 / Terminated / root cause: manufacturing process
Respironics, Inc. began a recall of Non-continuous ventilator accessories under the following brand names: a) REMStar (Tower) Ultrafine Qty 1 model number 1036915, b) REMStar Auto M. w/Humid... on . The reason FDA records is "Foreign material: Glass fragments may present in the plastic bag material used to ship components". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1065-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Respironics, Inc.
- Product
- Non-continuous ventilator accessories under the following brand names: a) REMStar (Tower) Ultrafine Qty 1 model number 1036915, b) REMStar Auto M. w/Humid. w/Smartcard. US model number DS500HS; c) REMStar Auto M. w/Smartcard A FLEX US model number DS510S; d) REMStar Auto M. w/Smartcard, CNDA model number CA501S; e) REMStar Auto M w/Smartcard Intl model number IN501S; f) REMStar Auto M w/Smartcard US model number DS500S; g) REMStar Auto w/C-FLEX CANADA model number CA1017454; and h) REMStar Auto w/C-FLEX Domestic Core Pkg model number 1017476. Filter, mask, and/or accessories for sleep therapy and ventilation devices, Respironics, Inc., Murrysville, PA 15668.
- Product code
- BZD Ventilator, Non-Continuous (Respirator)
- Reason for recall
- Foreign material: Glass fragments may present in the plastic bag material used to ship components.
- Root cause tag
- manufacturing process
- FDA root cause
- Environmental control
- Action
- The recalling firm issued a Recall Notification dated 10/18/07 to its customers. Replacement products were to be sent.
- Quantity in commerce
- 20950 units
- Distribution
- The products were shipped nationwide to medical facilities.
- Product codes and lots
- Lot Numbers 071010, 071011, and 071012.