Philips Medical Systems, GEMINI GXL 6, 10, & 16 slice PET/CT Systems, diagnostic imaging systems for fixed or...
Philips Medical Systems (Cleveland) Inc began a recall of Philips Medical Systems, GEMINI GXL 6, 10, & 16 slice PET/CT Systems, diagnostic imaging systems for fixed or mobile installations that combine Positron... on . The reason FDA records is "Artifacts: Artifacts may appear in some images from head scans on the CT subsystem of the GEMINI PET/CT system. Reading images with these artifacts may lead to misdiagnosis or improper treatment". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0462-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips Medical Systems, GEMINI GXL 6, 10, & 16 slice PET/CT Systems, diagnostic imaging systems for fixed or mobile installations that combine Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems, Philips Medical Systems, Cleveland, OH 44143
- Product code
- KPS System, Tomography, Computed, Emission
- Reason for recall
- Artifacts: Artifacts may appear in some images from head scans on the CT subsystem of the GEMINI PET/CT system. Reading images with these artifacts may lead to misdiagnosis or improper treatment.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Philips Medical notified domestic customers via Product Safety Notification dated May 30, 2007. The product safety notification includes a Technical Bulletin notifying customers of the issue and how to avoid it. This information will be used until the software correction becomes available. The software correction will be released in October 2008. Once software is available, the firm will begin correcting the field.
- Quantity in commerce
- 126
- Distribution
- Worldwide Distribution - USA, Australia, Belgium, Canada, China, Denmark, England, Finland, France, Germany, Hong Kong, Hungary, India, Ireland, Italy, Japan, Jordan, Korea, Netherlands, Poland, Portugal, Saudi Arabia, Slovenia, Spain, Sweden, Taiwan, Turkey, and Venezuela.
- Product codes and lots
- Gemini GXL Product Numbers/Manufacturer Numbers: GXL 16 slice: 4535 679 71891, GXL 10 slice: 4535 679 75681, GXL 6 slice: 4535 679 75691. Catalog Numbers: GXL 16 slice: 882410, GXL 10 slice: 882400, GXL 6 slice: 882390. Serial Numbers: 4034-4159.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
