Solar HA Humeral Stem with Plasma Coating 7mm to 17 mm Product Catalog Number: 5351-4307 to 5351-4317. For...
Stryker Howmedica Osteonics Corp. began a recall of Solar HA Humeral Stem with Plasma Coating 7mm to 17 mm Product Catalog Number: 5351-4307 to 5351-4317. For use as a Shoulder Replacement. Stryker Orthopaedics... on . The reason FDA records is "Delamination: Loosening of the titanium plasma coating on the product was observed on twelve non-distributed units". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1236-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Solar HA Humeral Stem with Plasma Coating 7mm to 17 mm Product Catalog Number: 5351-4307 to 5351-4317. For use as a Shoulder Replacement. Stryker Orthopaedics Made in USA
- Product code
- KWS Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented
- Reason for recall
- Delamination: Loosening of the titanium plasma coating on the product was observed on twelve non-distributed units.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On 11/28/2007, Stryker Orthopaedics sent, via FedEx, Urgent Product Recall notification letters to distributors, hospitals, surgeons and hospitals' Chiefs of Orthopaedics informing them of the firm's voluntary recall of the affected product. They were instructed to discontinue use of the product and to have the units returned to the firm. Distributors were further instructed to notify their customers using the sub-recall letter provided by the firm. The letter to surgeons also informed them to monitor their patients who had received an implant of the product.
- Quantity in commerce
- 915 units
- Distribution
- Nationwide - USA
- Product codes and lots
- All Lot Codes for the following: Catalog # 5351-4307, 7mm; 5351-4308, 8mm; 5351-4309, 9mm; 5351-4310, 10mm; 5351-4311, 11mm; 5351-4312, 12mm; 5351-4313, 13mm; 5351-4314, 14mm; 5351-4315, 15mm; 5351-4316, 16mm; and 5351-4317, 17mm.