Z-0918-2008

Stryker Modular Replacement System Curved Cemented Stem; MRS Small Curved Femoral Stem 8x102. Catalog Number...

Howmedica Osteonics Corp / initiated 2007-11-15 / Terminated / root cause: manufacturing process

Stryker Howmedica Osteonics Corp. began a recall of Stryker Modular Replacement System Curved Cemented Stem; MRS Small Curved Femoral Stem 8x102. Catalog Number 6485-3-308; Stryker Orthopaedics Mahwah, NJ 07430 on . The reason FDA records is "Curvature Discrepancy --The radius of curvature may be out of specification to varying degrees causing a discrepency in the stem curvature between trials and implants". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0918-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Stryker Modular Replacement System Curved Cemented Stem; MRS Small Curved Femoral Stem 8x102. Catalog Number 6485-3-308; Stryker Orthopaedics Mahwah, NJ 07430
Product code
JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
Reason for recall
Curvature Discrepancy --The radius of curvature may be out of specification to varying degrees causing a discrepency in the stem curvature between trials and implants.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
On November 26, 2007, Stryker Sent Urgent Product Recall notification letters to Stryker Agencies/Branches and hospitals (OR Supervisor) by Federal Express. The notification requested that all inventory be examined and returned. Returning instructions were provided in the letter. Questions concerning this recall are being addressed by contacting the Regulatory Compliance Hotline at (201) 831-6633.
Quantity in commerce
803 total, all sizes
Distribution
Nationwide Distribution
Product codes and lots
Catalog Number 6485-3-308;  Lot Codes: LCDBP;TEC024A;TEC024B;TEC080A;TEC080C.