Stryker Modular Replacement System Curved Cemented Stem; MRS 17x127 Curved Femoral Stem. Catalog Number...
Stryker Howmedica Osteonics Corp. began a recall of Stryker Modular Replacement System Curved Cemented Stem; MRS 17x127 Curved Femoral Stem. Catalog Number: 6485-3-717; Stryker Orthopaedics Mahwah, NJ 07430 on . The reason FDA records is "Curvature Discrepancy -- The radius of curvature may be out of specification to varying degrees causing a discrepency in the stem curvature between trials and implants". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0927-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Stryker Modular Replacement System Curved Cemented Stem; MRS 17x127 Curved Femoral Stem. Catalog Number: 6485-3-717; Stryker Orthopaedics Mahwah, NJ 07430
- Product code
- JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
- Reason for recall
- Curvature Discrepancy -- The radius of curvature may be out of specification to varying degrees causing a discrepency in the stem curvature between trials and implants.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- On November 26, 2007, Stryker Sent Urgent Product Recall notification letters to Stryker Agencies/Branches and hospitals (OR Supervisor) by Federal Express. The notification requested that all inventory be examined and returned. Returning instructions were provided in the letter. Questions concerning this recall are being addressed by contacting the Regulatory Compliance Hotline at (201) 831-6633.
- Quantity in commerce
- 803 total, all sizes
- Distribution
- Nationwide Distribution
- Product codes and lots
- Catalog Number: 6485-3-717; Lot Codes: LCDCH, TEC029A, TEC057A, TEC089B, TEC089C, TEC089D, TEC089E, TEC089E2, TEC089F, TEC089G, TEC089G1, TEC089H1, TEC089I1, TEC144, TEC145, TE 259, TEC301.