Z-1160-2008
Access Ultrasensitive Insulin Assay Kit, Part Number 33410. The product is a paramagnetic particle...
Siemens Healthcare Diagnostics Inc. / initiated 2007-09-19 / Terminated
Beckman Coulter Inc began a recall of Access Ultrasensitive Insulin Assay Kit, Part Number 33410. The product is a paramagnetic particle, chemiluminescent immunoassay for the quantitative... on . The reason FDA records is "False negative results when used to test serum samples (as compared to plasma samples)". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1160-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Access Ultrasensitive Insulin Assay Kit, Part Number 33410. The product is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of insulin levels in human serum and plasma (EDTA) using the Access Immunoassay Systems. Beckman Coulter, Inc., Chaska, MN 55318.
- Product code
- CFP Radioimmunoassay, Immunoreactive Insulin
- Reason for recall
- False negative results when used to test serum samples (as compared to plasma samples).
- FDA root cause
- Other
- Action
- On September 19, 2007, Beckman Coulter mailed to its consignees an Urgent Product Corrective Action Letter informing them that the affected lots of the product were giving false negative results when used to test serum samples (as compared to plasma samples). The letter instructed consignees to discontinue using the affected lots of the product, and to re-test using an EDTA plasma sample if there was any discordance between the results of the serum sample (if it was obtained using an affected lot of the product) as compared with the clinical presentaion of the patient and the results of other diagnostic tests. The letter also instructed consignees to return, within 10 days, the Response Sheet .
- Quantity in commerce
- 8,205 units in the US; 332 in Canada
- Distribution
- Worldwide: including USA and Canada.
- Product codes and lots
- Part Number 33410, Lot Numbers: 714362 (Exp dt. 3/31/09), 716129 (Exp dt. 6/30/09), 718258 (Exp dt. 7/31/09), 519096 (Exp dt. 10/31/07), 521487 (Exp dt. 2/28/08), 616274 (Exp dt. 7/31/08), 617890 (Exp dt. 7/31/08), 617891 (Exp dt. 7/31/08) and 620204 (Exp dt. 12/31/08)