Z-1137-2008

MultiDiagnost Eleva Flat Detector X-Ray System; Part Number: SW 9896 001 43361; Philips Medical Systems...

Philips North America LLC / initiated 2007-11-29 / Terminated / root cause: software

Philips Medical Systems North America Co. Phillips began a recall of MultiDiagnost Eleva Flat Detector X-Ray System; Part Number: SW 9896 001 43361; Philips Medical Systems, Bothell, WA, 98021 on . The reason FDA records is "Delay in Imaging: MultiDiagnost Eleva with Flat Detector x-ray system cannot perform fluoroscopy procedures while certain image file housekeeping operations are performed. Fluoroscopy cannot be performed and new...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1137-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
MultiDiagnost Eleva Flat Detector X-Ray System; Part Number: SW 9896 001 43361; Philips Medical Systems, Bothell, WA, 98021
Product code
JAA System, X-Ray, Fluoroscopic, Image-Intensified
Reason for recall
Delay in Imaging: MultiDiagnost Eleva with Flat Detector x-ray system cannot perform fluoroscopy procedures while certain image file housekeeping operations are performed. Fluoroscopy cannot be performed and new exposures are not able to be acquired until the housekeeping operations are complete which may take several minutes. The loss of system availability could occur during an interventional
Root cause tag
software
FDA root cause
Software design
Action
On 11/29/07 Philips Medical Systems sent a letter dated 11/28/07 to customers stating that if users perform certain image file housekeeping operations, fluoroscopy cannot be performed and new exposures cannot be acquired until those operations are complete. It also instructs that the unavailability of fluoroscopy and exposure can be avoided by not pressing the "Delete All Images" button until any pending Autopush operation is complete. FCO 70800067 will be issued to service representatives to install new software.
Quantity in commerce
9 units were distributed to hospitals in the US, 6 were installed, 3 were not installed or in use.
Distribution
USA Distribution - CO, MN, NC, NJ and TX.
Product codes and lots
Sysco Software Version 2, release 4.1.1