BD Vacutainer Push Button Blood collection Sets with Pre-Attached Holder; Catalog Number: 368656; 23 g X12 "...
Becton Dickinson & Company began a recall of BD Vacutainer Push Button Blood collection Sets with Pre-Attached Holder; Catalog Number: 368656; 23 g X12 " PBBCS w/PAH (Preattached), BD Diagnostics... on . The reason FDA records is "Safety Mechanism Failure: the safety mechanism of the BD Vacutainer Push Button Collection Set with Pre-attached holder cannot be activated; thus the needle cannot retract into the rear barrel". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0590-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BD Vacutainer Push Button Blood collection Sets with Pre-Attached Holder; Catalog Number: 368656; 23 g X12 " PBBCS w/PAH (Preattached), BD Diagnostics, Franklin Lakes, NJ
- Product code
- JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
- Reason for recall
- Safety Mechanism Failure: the safety mechanism of the BD Vacutainer Push Button Collection Set with Pre-attached holder cannot be activated; thus the needle cannot retract into the rear barrel.
- Root cause tag
- manufacturing process
- FDA root cause
- Material/Component Contamination
- Action
- BD sent Recall notification letters, Urgent Class II Product Recall, to customers and distributors on 12/10/07 through a third party, West Inc., Appleton, Wi. Mailings were sent return receipt requested. Letters requested immediate inventory search and product return. Contact number for questions was given as 1-800-631-0174
- Quantity in commerce
- 769,200 USA; 1,300 Canada
- Distribution
- Worldwide-USA and Canada
- Product codes and lots
- Lot #'s: 6090251R, 6094839, 6111757R, 6121669, 6121676, 6138669, 6146186R, 6152710, 6151443, 6160625, 6164839, 6171015, 6174156, 6178381, 7018205, 7025868, 7026017, 7033728, 7093399, 7093400, 7093403, 7106969, 7106970, 7106971, 7106972, 7113647, 7113648, 7124156, 7124157, 7131786, 7141678, 7141679, 7141680, 7150272, 7162775, 7167155, 7169403, 7172215, 7177261, and 7208536.