Z-0624-2009

Medtronic O-Arm Intraoperative Imaging System Catalog Number: BI-700-00027

Medtronic Navigation, Inc / initiated 2007-11-27 / Terminated / root cause: software

Medtronic Navigation, Inc began a recall of Medtronic O-Arm Intraoperative Imaging System Catalog Number: BI-700-00027 on . The reason FDA records is "Navigation accuracy problem: Accuracy may fall outside of specification when used in combination with the Medtronic Synergy Experience StealthStation System Spine Software with the O-Arm Intraoperative Imaging System...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0624-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
Medtronic O-Arm Intraoperative Imaging System Catalog Number: BI-700-00027
Product code
IZL System, X-Ray, Mobile
Reason for recall
Navigation accuracy problem: Accuracy may fall outside of specification when used in combination with the Medtronic Synergy Experience StealthStation System Spine Software with the O-Arm Intraoperative Imaging System and tracker.
Root cause tag
software
FDA root cause
Software design
Action
Medtronic Navigation notified Field representatives on 11/27/07 to perform site visits to implement FCA 2007-005 - O-Arm Intraoperative Imaging System Accuracy, and deliver the Customer Letter notification. Confirmation of the visit was documented.
Quantity in commerce
54 units
Distribution
Worldwide Distribution: USA, Canada, Czech Republic, Denmark, Germany, Netherlands, S. Korea, Poland, South Africa, and Singapore
Product codes and lots
Software versions prior to version 3.0.2 S/N: 102-155