Z-0475-2008
Hipstar TMZF Cementless Hip Stem Protocol, Product Number: LSP60
Howmedica Osteonics Corp / initiated 2007-09-28 / Terminated / root cause: labeling and UDI
Stryker Howmedica Osteonics Corp. began a recall of Hipstar TMZF Cementless Hip Stem Protocol, Product Number: LSP60 on . The reason FDA records is "Mis-labeling: An error on page 4 of the Hipstar protocol was found. In the table for compatible Heads, the 40 mm diameter heads with a Neck Length of +12 should have been omitted". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0475-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Hipstar TMZF Cementless Hip Stem Protocol, Product Number: LSP60.
- Product code
- LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Reason for recall
- Mis-labeling: An error on page 4 of the Hipstar protocol was found. In the table for compatible Heads, the 40 mm diameter heads with a Neck Length of +12 should have been omitted.
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- Urgent Product Correction Letters, dated 10/17/2007 were sent on 10/17/07by Federal Express to all Stryker Branches/Agencies who had received the protocol. The branches were instructed to quarantine and destroy the protocols.
- Quantity in commerce
- 192 protocols
- Distribution
- Nationwide:The protocol was distributed to 9 Stryker branches/agencies in NJ, MD, IN, MI, MN, IL and TX.
- Product codes and lots
- Literature # LPS60 (Lot code n/a)
FIRST ARTICLE / WEEKLY