Roche COBAS 6000 Clinical Chemistry analyzer; Catalog Number: 04745914001, Roche Diagnostics Corporation...

Roche Diagnostics GmbH / initiated 2007-12-04 / Terminated / root cause: software

Roche Diagnostics Corp. began a recall of Roche COBAS 6000 Clinical Chemistry analyzer; Catalog Number: 04745914001, Roche Diagnostics Corporation, Indianapolis, IN 46250 on . The reason FDA records is "Incorrect results: CRP assays may be calculated using incorrect calibration parameters which could result in falsely high or falsely low patient results being reported". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0576-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche COBAS 6000 Clinical Chemistry analyzer; Catalog Number: 04745914001, Roche Diagnostics Corporation, Indianapolis, IN 46250.
Product code
JJY Multi-Analyte Controls, All Kinds (Assayed)
Reason for recall
Incorrect results: CRP assays may be calculated using incorrect calibration parameters which could result in falsely high or falsely low patient results being reported.
Root cause tag
software
FDA root cause
Software design
Action
Roche Diagnostics contacted the consignee by phone on 12/4/07 and by follow-up letter dated 12/4/07. The user was instructed to discontinue use of two CRP assays on this analyzer. The letter described a workaround and that a new revision of software, due out first quarter 2008, will fix this.
Quantity in commerce
1
Distribution
USA Distribution in the state of Nebraska.
Product codes and lots
All cce and cc system configurations with core unit software version 02-03.