Z-0705-2008
Biomet RINGLOC + Instrumentation Impactor Plate, Size 28; REF/Part#: 31-400628, Biomet Orthopedics, Inc...
Zimmer GmbH / initiated 2007-10-18 / Terminated / root cause: manufacturing process
Biomet, Inc. began a recall of Biomet RINGLOC + Instrumentation Impactor Plate, Size 28; REF/Part#: 31-400628, Biomet Orthopedics, Inc., Warsaw, IN on . The reason FDA records is "The instrument will not lock onto the impactor handle; the device was distributed without assembly of the Bal Seal component". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0705-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Biomet RINGLOC + Instrumentation Impactor Plate, Size 28; REF/Part#: 31-400628, Biomet Orthopedics, Inc., Warsaw, IN.
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- The instrument will not lock onto the impactor handle; the device was distributed without assembly of the Bal Seal component.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Distributors were notified of the device defect by letter (Urgent Field Correction Notice) dated 10/18/07, and provided with assembly instructions and parts, and instructed to make corrections at consignee locations.
- Quantity in commerce
- 141 total for all recalls Z-0700-0705-2008
- Distribution
- Worldwide-USA, Germany, The Netherlands, Sweden and The United Kingdom.
- Product codes and lots
- Lot Number: 038730.