Z-0705-2008

Biomet RINGLOC + Instrumentation Impactor Plate, Size 28; REF/Part#: 31-400628, Biomet Orthopedics, Inc...

Zimmer GmbH / initiated 2007-10-18 / Terminated / root cause: manufacturing process

Biomet, Inc. began a recall of Biomet RINGLOC + Instrumentation Impactor Plate, Size 28; REF/Part#: 31-400628, Biomet Orthopedics, Inc., Warsaw, IN on . The reason FDA records is "The instrument will not lock onto the impactor handle; the device was distributed without assembly of the Bal Seal component". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0705-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Biomet RINGLOC + Instrumentation Impactor Plate, Size 28; REF/Part#: 31-400628, Biomet Orthopedics, Inc., Warsaw, IN.
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
The instrument will not lock onto the impactor handle; the device was distributed without assembly of the Bal Seal component.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Distributors were notified of the device defect by letter (Urgent Field Correction Notice) dated 10/18/07, and provided with assembly instructions and parts, and instructed to make corrections at consignee locations.
Quantity in commerce
141 total for all recalls Z-0700-0705-2008
Distribution
Worldwide-USA, Germany, The Netherlands, Sweden and The United Kingdom.
Product codes and lots
Lot Number: 038730.