Z-1114-2008

Pointe Scientific Liquid ALT (SGPT) reagent set for the quantitative determination of alanine...

Horiba Instruments Incorporated / initiated 2007-12-14 / Terminated / root cause: manufacturing process

Pointe Scientific, Inc. began a recall of Pointe Scientific Liquid ALT (SGPT) reagent set for the quantitative determination of alanine aminotransferase in serum, 610 ml, Catalog no. HA926-610, in... on . The reason FDA records is "Failure of the reagent to produce test results. The R1 reagent may be contaminated with microorganisms". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1114-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Horiba Instruments Incorporated
Product
Pointe Scientific Liquid ALT (SGPT) reagent set for the quantitative determination of alanine aminotransferase in serum, 610 ml, Catalog no. HA926-610, in vitro diagnostic, Pointe Scientific, Inc.; Canton, MI 48188
Product code
CKA Nadh Oxidation/Nad Reduction, Alt/Sgpt
Reason for recall
Failure of the reagent to produce test results. The R1 reagent may be contaminated with microorganisms.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Pointe Scientific notified consignees via Urgent Medical Device Recall letters dated 12/14/07, instructed to dispose of the product and to send back a reply form to request replacement product.
Quantity in commerce
955 for all products.
Distribution
Worldwide Distribution - USA, Bolivia, Canada, Chile, China, Cyprus, Greece, Indonesia, Israel, Korea, Mexico, Pakistan, Panama, Peru, Philippines, Republic of Yemen, Russia, South Africa, Trinidad, Ukraine and West Indies.
Product codes and lots
Lots 628302-004 and 704601-113.