Z-1102-2008
Seradyn QMS Vancomycin reagents, Part Number: 0373589. Seradyn, Inc, Indianapolis, IN
Seradyn, Inc. / initiated 2007-12-20 / Terminated / root cause: design
Seradyn, Inc. began a recall of Seradyn QMS Vancomycin reagents, Part Number: 0373589. Seradyn, Inc, Indianapolis, IN on . The reason FDA records is "False negative test results due to Interfering substances (heterophile antibodies) in the patient's blood may produce erroneously low results". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1102-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Seradyn, Inc.
- Product
- Seradyn QMS Vancomycin reagents, Part Number: 0373589. Seradyn, Inc, Indianapolis, IN.
- Product code
- LEH Radioimmunoassay, Vancomycin
- Reason for recall
- False negative test results due to Interfering substances (heterophile antibodies) in the patient's blood may produce erroneously low results.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Consignees were notified of the problem via Urgent Medical Device Correction letter dated 12/20/07. The letter provided information on the issues concerning the recall. The firm will revise the package insert to address the potential occurrence of false negative bias. If inconsistencies with Vancomycin level exist, confirm with an alternative test method. For further information please contact 1-800-428-4072.
- Quantity in commerce
- 1,819 kits plus bulk material for 1,863 kits.
- Distribution
- Worldwide Distribution --- USA including states of California and New York, and countries of Australia, Belgium, Canada, Finland, Germany, Ireland, Japan, Spain and United Kingdom.
- Product codes and lots
- All lots (lots 501930, 502063, 502074, 600432, 601166, 601207, 602556, 603076, 603801, 603312, 700657 and 703027).