Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0579-2008

TENET Medical Engineering Wrist Stabilization Kit, Large; REF (Part) #7693; Sterile; manufactured by TENET...

Tenet Medical Engineering / initiated 2007-11-23 / Terminated / root cause: packaging

Tenet Medical Engineering Inc began a recall of TENET Medical Engineering Wrist Stabilization Kit, Large; REF (Part) #7693; Sterile; manufactured by TENET Medical Engineering, Calgary, Alberta, Canada T2Z 4M3 on . The reason FDA records is "Sterility compromised: Sterility seals on some of the firm's wrist stabilization kit pouches may be compromised". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0579-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Tenet Medical Engineering
Product
TENET Medical Engineering Wrist Stabilization Kit, Large; REF (Part) #7693; Sterile; manufactured by TENET Medical Engineering, Calgary, Alberta, Canada T2Z 4M3.
Product code
JEB Table, Surgical With Orthopedic Accessories, Manual
Reason for recall
Sterility compromised: Sterility seals on some of the firm's wrist stabilization kit pouches may be compromised.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
TENET notified US distributor consignee via Urgent Product Recall Notification letter sent FedEx on 11/23/07. Consignee asked to return product and to contact their consignees to also return product.
Quantity in commerce
160 units.
Distribution
Nationwide Distribution.
Product codes and lots
Lot Number: 20060605.