Siemens Mevatron KDS-2 Medical Charged-Particle Radiation Therapy System; (Interlock #51 Jaw Field Position)...
Siemens Medical Solutions USA, Inc began a recall of Siemens Mevatron KDS-2 Medical Charged-Particle Radiation Therapy System; (Interlock #51 Jaw Field Position) Model Number 9822693 and 9411588, Siemens Medical... on . The reason FDA records is "Incorrect field size -- Products manufactured prior to 1999 may experience jaw field size errors during interlock resulting in potential mistreatment if a light field check is not performed. This problem is present in...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0680-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens Mevatron KDS-2 Medical Charged-Particle Radiation Therapy System; (Interlock #51 Jaw Field Position) Model Number 9822693 and 9411588, Siemens Medical Solutions, Concord, CA 94520
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Incorrect field size -- Products manufactured prior to 1999 may experience jaw field size errors during interlock resulting in potential mistreatment if a light field check is not performed. This problem is present in machines with serial numbers less than 3094, as they do not contain redundant sensors for jaw position.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm has sent consignees a Customer Safety Advisory Notice on August 7, 2007 informing them that a misplacement of the X and/or Y jaws can occur if interlock #51 Jaw Field Position is asserted and then cleared with the left-hand key rest. The firm recommends that if an interlock issue is encountered tha treatment of the patient be discontined until this issue is resolved.
- Quantity in commerce
- 35 units
- Distribution
- Worldwide Distribution
- Product codes and lots
- Serial Numbers: 2108, 2183, 2244, 2251, 2263, 2272, 2284, 2289, 2335, 2357, 2084, 2168, 2264, 2304, 2394, 2398, 2023, 2055, 2097, 2106, 2122, 2158, 2208, 2222, 2242, 2269, 2276, 2324, 2336, 2339, 2353, 2358, 2372, 2196, 2229