Siemens Mevatron MDX-2 Medical Charged-Particle Radiation Therapy System; (Interlock #51 Jaw Field Position)...
Siemens Medical Solutions USA, Inc began a recall of Siemens Mevatron MDX-2 Medical Charged-Particle Radiation Therapy System; (Interlock #51 Jaw Field Position) Model number 9401746, Siemens Medical Solutions... on . The reason FDA records is "Incorrect field size -- Products manufactured prior to 1999 may experience jaw field size errors during interlock resulting in potential mistreatment if a light field check is not performed. This problem is present in...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0682-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens Mevatron MDX-2 Medical Charged-Particle Radiation Therapy System; (Interlock #51 Jaw Field Position) Model number 9401746, Siemens Medical Solutions, Concord, CA 94520
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Incorrect field size -- Products manufactured prior to 1999 may experience jaw field size errors during interlock resulting in potential mistreatment if a light field check is not performed. This problem is present in machines with serial numbers less than 3094, as they do not contain redundant sensors for jaw position.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm has sent consignees a Customer Safety Advisory Notice on August 7, 2007 informing them that a misplacement of the X and/or Y jaws can occur if interlock #51 Jaw Field Position is asserted and then cleared with the left-hand key rest. The firm recommends that if an interlock issue is encountered tha treatment of the patient be discontined until this issue is resolved.
- Quantity in commerce
- 2 units
- Distribution
- Worldwide Distribution
- Product codes and lots
- Serial Numbers: 2125 and 2137