Siemens Mevatron K-2 Medical Charged-Particle Radiation Therapy System; (Interlock #51 Jaw Field Position)...
Siemens Medical Solutions USA, Inc began a recall of Siemens Mevatron K-2 Medical Charged-Particle Radiation Therapy System; (Interlock #51 Jaw Field Position) Model Number: 1940753, Siemens Medical Solutions... on . The reason FDA records is "Incorrect field size -- Products manufactured prior to 1999 may experience jaw field size errors during interlock resulting in potential mistreatment if a light field check is not performed. This problem is present in...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0685-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens Mevatron K-2 Medical Charged-Particle Radiation Therapy System; (Interlock #51 Jaw Field Position) Model Number: 1940753, Siemens Medical Solutions, Concord, CA 94520
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Incorrect field size -- Products manufactured prior to 1999 may experience jaw field size errors during interlock resulting in potential mistreatment if a light field check is not performed. This problem is present in machines with serial numbers less than 3094, as they do not contain redundant sensors for jaw position.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm has sent consignees a Customer Safety Advisory Notice on August 7, 2007 informing them that a misplacement of the X and/or Y jaws can occur if interlock #51 Jaw Field Position is asserted and then cleared with the left-hand key rest. The firm recommends that if an interlock issue is encountered tha treatment of the patient be discontined until this issue is resolved.
- Quantity in commerce
- 59 units
- Distribution
- Worldwide Distribution
- Product codes and lots
- Serial Numbers: 2568, 2698, 2038, 2168, 2448, 2451, 2452, 2459, 2466, 2498, 2505, 2525, 2531, 2547, 2561, 2575, 2589, 2592, 2616, 2648, 2663, 2667, 2681, 2686, 2726, 2729, 2751, 2769, 2777, 2781, 2801, 2807, 1798, 2213, 2216, 2218, 2571, 2582, 2602, 2607, 2612, 2636, 2731, 2877, 2951, 2952, 3021, 3029, 3041, 1985, 2145, 2229, 2297, 2353, 2509, 2601, 2754, 2776, and 2394