Z-0652-2008
Stryker Orthopaedics; UHR Universal Bipolar Component; 26 x 61 mm, Hip Prosthesis Component, Stryker...
Howmedica Osteonics Corp / initiated 2007-10-26 / Terminated / root cause: manufacturing process
Stryker Howmedica Osteonics Corp. began a recall of Stryker Orthopaedics; UHR Universal Bipolar Component; 26 x 61 mm, Hip Prosthesis Component, Stryker Howmedica Osteonics Corp., Mahwah, NJ 07430 on . The reason FDA records is "Incorrect part: Device may have an incorrect locking ring assembled with the UHR bipolar head". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0652-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Stryker Orthopaedics; UHR Universal Bipolar Component; 26 x 61 mm, Hip Prosthesis Component, Stryker Howmedica Osteonics Corp., Mahwah, NJ 07430
- Product code
- KWL Prosthesis, Hip, Hemi-, Femoral, Metal
- Reason for recall
- Incorrect part: Device may have an incorrect locking ring assembled with the UHR bipolar head.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Stryker sent Urgent Product Hold letters to US branches/agencies and hospitals on October 26, 2007 advising them to quarantine the product from these lots, and Urgent Product Recall letters on November 2, 2007 advising them to return the devices to Stryker. International distributors were sent an advisory on November 6, 2007.
- Quantity in commerce
- 360 total
- Distribution
- Worldwide Distribution - USA, Canada, China, India, Australia, Korea, Singapore, Taiwan, Sweden, Germany, Netherlands, France, Spain, Italy, Poland, South Africa, and England.
- Product codes and lots
- Ref No: Lot Code: UH1-61-26 EDWMND