Siemens Coherence AG Therapist, Part No: 5863506, medical charged-particle radiation therapy system with...
Siemens Medical Solutions USA, Inc began a recall of Siemens Coherence AG Therapist, Part No: 5863506, medical charged-particle radiation therapy system with software version 2.1. Siemens AG, Concord, CA on . The reason FDA records is "Image Orientation Incorrect (Image Shifts) - 1) if the reference image has been calibrated for centering and the reference image is used for patient positioning, the image will shift when the positioning tools are used...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0971-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens Coherence AG Therapist, Part No: 5863506, medical charged-particle radiation therapy system with software version 2.1. Siemens AG, Concord, CA.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Image Orientation Incorrect (Image Shifts) - 1) if the reference image has been calibrated for centering and the reference image is used for patient positioning, the image will shift when the positioning tools are used. 2) This problem occurs only when non-square reference images are used, such as when images are cropped. At this point, the images will refresh, and the center of the reference im
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On 10/03/2007, the firm sent Consignees a Customer Safety Advisory Notice informing them of the two potential issues with the display of reference images used for position correction. Consignees will be informed when a solution to this issue has been validated.
- Quantity in commerce
- 7 units
- Distribution
- Worldwide Distribution-USA including 48 consignees in the US and 39 outside the US.
- Product codes and lots
- Serial Numbers: 84601, 84602, 85703, 85714, 85718, 85720, and 00E0815479E1