Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0973-2008

Siemens Coherence Impression Therapist system, Part No: 7341410, medical charged-particle radiation therapy...

Siemens Medical Solutions USA, Inc. / initiated 2007-10-02 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Siemens Coherence Impression Therapist system, Part No: 7341410, medical charged-particle radiation therapy system with software version 2.1. Siemens AG... on . The reason FDA records is "Image Orientation Incorrect (Image Shifts) - 1) if the reference image has been calibrated for centering and the reference image is used for patient positioning, the image will shift when the positioning tools are used...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0973-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens Coherence Impression Therapist system, Part No: 7341410, medical charged-particle radiation therapy system with software version 2.1. Siemens AG, Concord, CA.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
Image Orientation Incorrect (Image Shifts) - 1) if the reference image has been calibrated for centering and the reference image is used for patient positioning, the image will shift when the positioning tools are used. 2) This problem occurs only when non-square reference images are used, such as when images are cropped. At this point, the images will refresh, and the center of the reference im
Root cause tag
software
FDA root cause
Software design
Action
On 10/03/2007, the firm sent Consignees a Customer Safety Advisory Notice informing them of the two potential issues with the display of reference images used for position correction. Consignees will be informed when a solution to this issue has been validated.
Quantity in commerce
16 units
Distribution
Worldwide Distribution-USA including 48 consignees in the US and 39 outside the US.
Product codes and lots
Serial Numbers: 2801, 10201, 14101, 22201, 56302, 57401, 62701, 79102, 93301, 98001, 00E0815462E6, 00E08154754E, 00E081548949, 102001, 1601, and 39901.