Bayer Contour TS blood glucose test strips packaged in packs of 10, 25, 50 and 100 strips; Bayer Healthcare...
Bayer Healthcare LLC began a recall of Bayer Contour TS blood glucose test strips packaged in packs of 10, 25, 50 and 100 strips; Bayer Healthcare, Mishawaka, IN. U.S. Product codes 1820, 1823, 1825... on . The reason FDA records is "Inaccurate Test Strips Results: Results in blood glucose readings with a positive bias are outside the product specifications. Patient test results may demonstrative results 5 percent to 17 percent higher". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0932-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medrad, Inc.
- Product
- Bayer Contour TS blood glucose test strips packaged in packs of 10, 25, 50 and 100 strips; Bayer Healthcare, Mishawaka, IN. U.S. Product codes 1820, 1823, 1825 and 9578; 1835 and 1836 (Korea); 1832, 1833 and 1837 (India); 1823 and 1825 (Mexico).
- Product code
- LFR Glucose Dehydrogenase, Glucose
- Reason for recall
- Inaccurate Test Strips Results: Results in blood glucose readings with a positive bias are outside the product specifications. Patient test results may demonstrative results 5 percent to 17 percent higher.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- The firm issued a Press Release on 12/21/07. A Test Strip Recall letter was also sent to the firm's consignees and health care providers instructing them that use of the product should cease and that health care providers contact the firm for replacements. Additional information can be found at www.bayerdiabeties.com or by contacting the Bayer at 1-800-348-8100.
- Quantity in commerce
- 97,000 bottles of test strips
- Distribution
- Worldwide Distribution --- including USA and countries of Austria, France, India, Korea, Mexico, Philippines and Turkey.
- Product codes and lots
- Test strip lots WK7DD3E31A, WK7DD3E31B, WK7DD3E31D, WK7DD3E51C, WK7DD3E51D, WK7ED3E02A, WK7ED3E04A, WK7ED3E51B, WK7ED3E52A, WK7ED3E52C, WK7ED3E52D, WK7ED3E52E, WK7ED3E53C, WK7ED3E53F, WK7FD3E01C, WK7FD3E01D, WK7FD3E01E, WK7FD3E04A, WK7FD3E31C, WK7FD3E31D, WK7FD3E31E, WK7FD3E31F, WK7FD3E31H, WK7FD3E32C, WK7FD3E34A, WK7FD3E34B, WK7FD3E51A, WK7FD3E51B, WK7GD3E01A, WK7GD3E02A, WK7GD3E05A, WK7GD3E51A and WK7GD3E51B.