Z-0664-2008

Tucker Embryo Catheter with URM Tip, Embryo Transfer Catheter, 20 cm, Order #: ftr120, Manufactured by...

Fertility Technology Resources, Inc. / initiated 2007-09-13 / Terminated / root cause: manufacturing process

Fertility Technology Resources, Inc. began a recall of Tucker Embryo Catheter with URM Tip, Embryo Transfer Catheter, 20 cm, Order #: ftr120, Manufactured by: Fertility Technology Resources, Inc., Marietta, GA 30062 on . The reason FDA records is "Expiration Date for sterility; The 5 year dating could not be validated/supported for sterility assurance". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0664-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Fertility Technology Resources, Inc.
Product
Tucker Embryo Catheter with URM Tip, Embryo Transfer Catheter, 20 cm, Order #: ftr120, Manufactured by: Fertility Technology Resources, Inc., Marietta, GA 30062
Product code
MQF Catheter, Assisted Reproduction
Reason for recall
Expiration Date for sterility; The 5 year dating could not be validated/supported for sterility assurance.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
All customers were contacted by telephone starting 9/13/2007. They were asked to return all Tucker family of catheters and all Morton IUI catheters that are on the shelves. A UPS call tag was sent to each customer to facilitate the return of the products. A follow up Urgent Device Recall letter was sent to all customers on 10/9/2007 reminding them to return all outstanding inventory when the UPS call tag arrives.
Quantity in commerce
not reported
Distribution
Worldwide: USA including states of CA, CO, FL, GA, LA, MD, ME, MI, MS, NC, NH, NJ, NY, OH, TN, and WA and The Netherlands.
Product codes and lots
All units shipped before August 17, 2007