Kodak DirectView Model DR 3000 System, Stationary X-ray System; Catalog #s 6551360, 6551386, 6552152...

Eastman Kodak Company / initiated 2007-12-21 / Terminated / root cause: software

Carestream Health, Inc. began a recall of Kodak DirectView Model DR 3000 System, Stationary X-ray System; Catalog #s 6551360, 6551386, 6552152, 6552160, 6551345, 6552319, 6552335. The units are labeled... on . The reason FDA records is "Unintended Positioner Movement - The U-Arm Positioner starts to move 1) without command from the operator and/or 2) can unexpectedly start movement if the system is powered down and re-energized after a collision has...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0731-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Eastman Kodak Company
Product
Kodak DirectView Model DR 3000 System, Stationary X-ray System; Catalog #s 6551360, 6551386, 6552152, 6552160, 6551345, 6552319, 6552335. The units are labeled under the following firms: EASTMAN KODAK COMPANY (now known as Carestream Heath Inc.) Rochester, NY 14608, or CARESTREAM HEALTH, INC., Rochester, NY 14608.
Product code
KPR System, X-Ray, Stationary
Reason for recall
Unintended Positioner Movement - The U-Arm Positioner starts to move 1) without command from the operator and/or 2) can unexpectedly start movement if the system is powered down and re-energized after a collision has occurred with the Quantum Table.
Root cause tag
software
FDA root cause
Software design
Action
On 12/21/07, Service Bulletins (dated December 2007) were sent to consignees informing them of design modification, and providing modification instructions. The firm followed up with telephone calls to the direct user sites to schedule appointments to visit each site, in order to make the necessary modifications. The firm will be installing Safety Adaptation Board and software into all the affected DR3000 systems.
Quantity in commerce
Domestic - 19 units
Distribution
Nationwide Distribution including states of NY, PA, MA, MI, IL, NC, GA, FL, KS, MO, TX, IA, and WA.
Product codes and lots
Serial Numbers: 30000032, 30001001, 30001004, 30001006, 30001016, 30001020, 30001024, 30001036, 30001038, 30001040, 30001039, 30001042, 30001048, 30001043, 30001049, 30001044, 30001045, 30001046, 30001021.