Kodak DirectView Model DR 3000 System, Stationary X-ray System; Catalog #s 6551360, 6551386, 6552152...
Carestream Health, Inc. began a recall of Kodak DirectView Model DR 3000 System, Stationary X-ray System; Catalog #s 6551360, 6551386, 6552152, 6552160, 6551345, 6552319, 6552335. The units are labeled... on . The reason FDA records is "Unintended Positioner Movement - The U-Arm Positioner starts to move 1) without command from the operator and/or 2) can unexpectedly start movement if the system is powered down and re-energized after a collision has...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0731-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Eastman Kodak Company
- Product
- Kodak DirectView Model DR 3000 System, Stationary X-ray System; Catalog #s 6551360, 6551386, 6552152, 6552160, 6551345, 6552319, 6552335. The units are labeled under the following firms: EASTMAN KODAK COMPANY (now known as Carestream Heath Inc.) Rochester, NY 14608, or CARESTREAM HEALTH, INC., Rochester, NY 14608.
- Product code
- KPR System, X-Ray, Stationary
- Reason for recall
- Unintended Positioner Movement - The U-Arm Positioner starts to move 1) without command from the operator and/or 2) can unexpectedly start movement if the system is powered down and re-energized after a collision has occurred with the Quantum Table.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On 12/21/07, Service Bulletins (dated December 2007) were sent to consignees informing them of design modification, and providing modification instructions. The firm followed up with telephone calls to the direct user sites to schedule appointments to visit each site, in order to make the necessary modifications. The firm will be installing Safety Adaptation Board and software into all the affected DR3000 systems.
- Quantity in commerce
- Domestic - 19 units
- Distribution
- Nationwide Distribution including states of NY, PA, MA, MI, IL, NC, GA, FL, KS, MO, TX, IA, and WA.
- Product codes and lots
- Serial Numbers: 30000032, 30001001, 30001004, 30001006, 30001016, 30001020, 30001024, 30001036, 30001038, 30001040, 30001039, 30001042, 30001048, 30001043, 30001049, 30001044, 30001045, 30001046, 30001021.