Z-1687-2008
Stryker Medical Epic II Critical Care Bed with Zoom Drive System, 115 V, Stryker Medical, Portage, MI.; Model...
Stryker Medical / initiated 2008-02-01 / Terminated / root cause: design
Stryker Medical Div. of Stryker Corporation began a recall of Stryker Medical Epic II Critical Care Bed with Zoom Drive System, 115 V, Stryker Medical, Portage, MI.; Model 2040 on . The reason FDA records is "The brakes may not have adequate holding power to lock the bed in place". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1687-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Medical
- Product
- Stryker Medical Epic II Critical Care Bed with Zoom Drive System, 115 V, Stryker Medical, Portage, MI.; Model 2040.
- Product code
- FNL Bed, Ac-Powered Adjustable Hospital
- Reason for recall
- The brakes may not have adequate holding power to lock the bed in place.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Consignees were notified via letter dated 2/1/08 and informed that upgrade kits along with full written and video installation instructions will be sent to correct the problem.
- Quantity in commerce
- 66,558 of all models.
- Distribution
- Worldwide, Argentina, Brazil, Canada, Chile, China, Greece, India Japan, Mexico, Poland, Singapore, South Africa, South Korea, Switzerland, Taiwan and United Kingdom.
- Product codes and lots
- Serial numbers 0409060516 through 0711096116.