Z-1220-2008
VECTRA GENISYS, Model No. 2761, Chattanooga Group, Hixson, TN 37343, Made in the USA
Chattanooga Corp Inc / initiated 2007-12-17 / Terminated / root cause: design
Chattanooga Corp Inc began a recall of VECTRA GENISYS, Model No. 2761, Chattanooga Group, Hixson, TN 37343, Made in the USA on . The reason FDA records is "Device malfunction: causing electric shock and burn". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1220-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Chattanooga Corp Inc
- Product
- VECTRA GENISYS, Model No. 2761, Chattanooga Group, Hixson, TN 37343, Made in the USA
- Product code
- MPH
- Reason for recall
- Device malfunction: causing electric shock and burn.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Beginning on 12/17/2007, Chattanooga Corp. sent two recall notification letters, via UPS mail, to its consignees informing them of the situation. The letters stated that the recall was to the user level and requested that the user perform the software upgrade, which will eliminate the possibility of shock and burn.
- Quantity in commerce
- 4047 units
- Distribution
- Worldwide Distribution - including USA
- Product codes and lots
- Serial numbers 1000 - 5047