Z-0738-2008

Optiflux 180NR Advanced Fresenius Polysulfone, Catalog No.: 0500318E, Fresenius Medical Care North America...

Fresenius Medical Care Renal Therapies Group, LLC / initiated 2007-12-03 / Terminated / root cause: manufacturing process

Fresenius Medical Care North America began a recall of Optiflux 180NR Advanced Fresenius Polysulfone, Catalog No.: 0500318E, Fresenius Medical Care North America, Waltham, MA 02451. (Hemodialyzer) on . The reason FDA records is "Leaking: the Dialyzer may leak at the header resulting in small amounts of blood loss". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0738-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Product
Optiflux 180NR Advanced Fresenius Polysulfone, Catalog No.: 0500318E, Fresenius Medical Care North America, Waltham, MA 02451. (Hemodialyzer)
Product code
KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
Reason for recall
Leaking: the Dialyzer may leak at the header resulting in small amounts of blood loss.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Consignees were notified by telephone and certified letter , Urgent Device Recall, on 12/03/2007. They were told to discontinue use and to return all unused affected dialyzers to Fresenius.
Quantity in commerce
2,791 cases (33,492 dialyzers)
Distribution
Nationwide including the states of AZ, CA, HI, IA, ID, IL, IN, KS, KY, MI, MN, MO, NE, NV, NY, OH, OR, PA, WA, and WI.
Product codes and lots
Lot # 7LU416