Z-0738-2008
Optiflux 180NR Advanced Fresenius Polysulfone, Catalog No.: 0500318E, Fresenius Medical Care North America...
Fresenius Medical Care Renal Therapies Group, LLC / initiated 2007-12-03 / Terminated / root cause: manufacturing process
Fresenius Medical Care North America began a recall of Optiflux 180NR Advanced Fresenius Polysulfone, Catalog No.: 0500318E, Fresenius Medical Care North America, Waltham, MA 02451. (Hemodialyzer) on . The reason FDA records is "Leaking: the Dialyzer may leak at the header resulting in small amounts of blood loss". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0738-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Fresenius Medical Care Renal Therapies Group, LLC
- Product
- Optiflux 180NR Advanced Fresenius Polysulfone, Catalog No.: 0500318E, Fresenius Medical Care North America, Waltham, MA 02451. (Hemodialyzer)
- Product code
- KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
- Reason for recall
- Leaking: the Dialyzer may leak at the header resulting in small amounts of blood loss.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Consignees were notified by telephone and certified letter , Urgent Device Recall, on 12/03/2007. They were told to discontinue use and to return all unused affected dialyzers to Fresenius.
- Quantity in commerce
- 2,791 cases (33,492 dialyzers)
- Distribution
- Nationwide including the states of AZ, CA, HI, IA, ID, IL, IN, KS, KY, MI, MN, MO, NE, NV, NY, OH, OR, PA, WA, and WI.
- Product codes and lots
- Lot # 7LU416