Z-0359-2009

Terumo Sarns Sternal Saw II System Power Unit, 220/240 V, Terumo Cardiovascular Systems Corp., Ann Arbor, MI...

Terumo Cardiovascular Systems Corporation / initiated 2008-02-14 / Terminated / root cause: manufacturing process

Terumo Cardiovascular Systems Corporation began a recall of Terumo Sarns Sternal Saw II System Power Unit, 220/240 V, Terumo Cardiovascular Systems Corp., Ann Arbor, MI; Catalog # 7084. Note: This product was not sold... on . The reason FDA records is "The drive power cable cannot be fully inserted into the power unit or can only be inserted with difficulty, resulting in the device not being operational". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0359-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo Sarns Sternal Saw II System Power Unit, 220/240 V, Terumo Cardiovascular Systems Corp., Ann Arbor, MI; Catalog # 7084. Note: This product was not sold in the U.S. Indicated for use in medial sternotomies. The Sarns Sternal Saw II system consists of a disposable blade, saw, flexible drive shaft, motor and foot control.
Product code
GFA Blade, Saw, General & Plastic Surgery, Surgical
Reason for recall
The drive power cable cannot be fully inserted into the power unit or can only be inserted with difficulty, resulting in the device not being operational.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Consignees were notified via Urgent Medical Device Removal letter dated 2/14/08 to not use the device if it exhibited the problem, provided a response card, and stated that replacement motors will be shipped to the person listed on the returned response card.
Quantity in commerce
69 total
Distribution
Worldwide Distribution --- including USA and countries of Belgium, China, Columbia, Malaysia, Philippines, United Arab Emirates, and Venezuela.
Product codes and lots
Serial numbers 1877 through 1925.