Z-1183-2008

Access Power Supply Assembly Sled Part Number: A48196/6845C; Component of: Access and Access 2 Immunoassay...

Siemens Healthcare Diagnostics Inc. / initiated 2007-10-17 / Terminated / root cause: manufacturing process

Beckman Coulter Inc began a recall of Access Power Supply Assembly Sled Part Number: A48196/6845C; Component of: Access and Access 2 Immunoassay Systems, SYNCHRON LXi 725 System, and UniCel DxC... on . The reason FDA records is "Electrical grounding failure: During manufacturing, internal testing found one instrument wlth an intermittent failure of the electrical grounding. The source of the problem was traced to inadequate soldering". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1183-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Access Power Supply Assembly Sled Part Number: A48196/6845C; Component of: Access and Access 2 Immunoassay Systems, SYNCHRON LXi 725 System, and UniCel DxC 600i System, for in vitro diagnostic use.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Electrical grounding failure: During manufacturing, internal testing found one instrument wlth an intermittent failure of the electrical grounding. The source of the problem was traced to inadequate soldering.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Beckman Coulter mailed a Product Corrective Action (PCA) letter on October 17, 2007 to all Access and Access 2 lmmunoassay Systems, SYNCHRON LXi 725 System, and UniCel DxC 600i System customers. The customers were informed that during manufacturing, internal testing found one instrument wlth an intermittent failure of the electrical grounding. The source of the problem was traced to inadequate soldering. The letter referenced a proactive initiative to inspect the power supply sled assemblies on all instruments referenced. A response form was included.
Quantity in commerce
3055 instruments in the US; 192 instruments in Canada
Distribution
Nationwide Distribution and Canada
Product codes and lots
Part Number A48196/6845C; Instrument part numbers 81600 and  81600N