Access Power Supply Assembly Sled Part Number: A48196/6845C; Component of: Access and Access 2 Immunoassay...
Beckman Coulter Inc began a recall of Access Power Supply Assembly Sled Part Number: A48196/6845C; Component of: Access and Access 2 Immunoassay Systems, SYNCHRON LXi 725 System, and UniCel DxC... on . The reason FDA records is "Electrical grounding failure: During manufacturing, internal testing found one instrument wlth an intermittent failure of the electrical grounding. The source of the problem was traced to inadequate soldering". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1183-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Access Power Supply Assembly Sled Part Number: A48196/6845C; Component of: Access and Access 2 Immunoassay Systems, SYNCHRON LXi 725 System, and UniCel DxC 600i System, for in vitro diagnostic use.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Electrical grounding failure: During manufacturing, internal testing found one instrument wlth an intermittent failure of the electrical grounding. The source of the problem was traced to inadequate soldering.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Beckman Coulter mailed a Product Corrective Action (PCA) letter on October 17, 2007 to all Access and Access 2 lmmunoassay Systems, SYNCHRON LXi 725 System, and UniCel DxC 600i System customers. The customers were informed that during manufacturing, internal testing found one instrument wlth an intermittent failure of the electrical grounding. The source of the problem was traced to inadequate soldering. The letter referenced a proactive initiative to inspect the power supply sled assemblies on all instruments referenced. A response form was included.
- Quantity in commerce
- 3055 instruments in the US; 192 instruments in Canada
- Distribution
- Nationwide Distribution and Canada
- Product codes and lots
- Part Number A48196/6845C; Instrument part numbers 81600 and 81600N