Siemens Syngo Imaging System with VB30A version of software. Model Number: 10014063, Siemans Medical...
Siemens Medical Solutions USA, Inc began a recall of Siemens Syngo Imaging System with VB30A version of software. Model Number: 10014063, Siemans Medical Solutions, Malvern, PA 19355 on . The reason FDA records is "Inconsistent data - A buffer overflow may occur between information system and picture system if the system experiences a high load. (e.g. RIS/PACS communication is unusally high)". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0732-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens Syngo Imaging System with VB30A version of software. Model Number: 10014063, Siemans Medical Solutions, Malvern, PA 19355
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Inconsistent data - A buffer overflow may occur between information system and picture system if the system experiences a high load. (e.g. RIS/PACS communication is unusally high)
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The recalling firm issued a Customer Safety Advisory, via certified mail. to the affected customers on 11/27/07. The letter informs the customer of the issue and also provides instruction to detect and avoid its occurrence. In addition, the firm is developing a software update to correct the issue, which will be installed on the affected systems when it is released.
- Quantity in commerce
- 10 units
- Distribution
- Nationwide Distribution, including states of AL, MO, NC, NY, OH, OR, TN, and TX.
- Product codes and lots
- Serial Numbers: 1005, 1065, 1076, 1112, 1202, 1242, 1269, 1278, 1279, and 1282.