Beckman Coulter UniCel Dxl 800 Access Immunoassay System; Part Number: 973100 With: UniCel Dxl 800...

Siemens Healthcare Diagnostics Inc. / initiated 2007-10-08 / Terminated / root cause: design

Beckman Coulter Inc began a recall of Beckman Coulter UniCel Dxl 800 Access Immunoassay System; Part Number: 973100 With: UniCel Dxl 800 Modification (MOD ID 10027) and/or UniCel Dxl 800... on . The reason FDA records is "Incorrect results: There is an increased variance in assay performance when the UniCel Dxl 800 Modification (MOD ID 10027) and/or UniCel Dxl 800 Modification (MOD ID 6733) have been implemented and there are room...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1184-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Beckman Coulter UniCel Dxl 800 Access Immunoassay System; Part Number: 973100 With: UniCel Dxl 800 Modification (MOD ID 10027) and/or UniCel Dxl 800 Modification (MOD ID 6733), for in vitro diagnostic use.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Incorrect results: There is an increased variance in assay performance when the UniCel Dxl 800 Modification (MOD ID 10027) and/or UniCel Dxl 800 Modification (MOD ID 6733) have been implemented and there are room fluctuations after assay calibration between the extremes of the 18C and 32C operating range.
Root cause tag
design
FDA root cause
Device Design
Action
Beckman Coulter mailed a Product Corrective Action (PCA) letter the week of October 8, 2007 to the customers who have the affected UniCel Dxl 800 Access lmmunoassay systems advising that there is an increased variance in assay performance when the UniCel Dxl 800 Modification (MOD ID 10027) and/or UniCel Dxl 800 Modification (MOD ID 6733) have been implemented and there are room fluctuations after assay calibration between the extremes of the 18C and 32C operating range. The letter further advises that Beckman Coulter will contact the customer to arrange a date to remove the modificatlons but until the modification is completed, the system should be operated in a room temperature of 20C to 30C. This can be acomplished by monitoring and recording room tamperature in the area of the instrument. Customers are also asked to review historical quality control records to ensure assay performance is within their laboratory specification and that patient results should be reviewed in context of other diagnostics tests and the clinical presentations of the patient. A response form is included.
Quantity in commerce
111 units in the US; 9 units in Canada
Distribution
Nationwide and Canada
Product codes and lots
Part Number 973100; Serial Numbers: 600469, 601097, 601098, 601144, 601182, 601160, 601188, 601071, 601117, 601114, 601137, 601059, 601039, 601099, 601088, 601176, 601184, 601178, 601025, 601186, 601110, 601063, 601002, 601029, 601133, 601106, 601215, 601159, 601077, 601202, 601206, 601079, 601051, 601179, 601130, 601058, 601031, 601212, 601173, 601028, 601078, 601094, 601210, 601189, 601228, 601043, 601092, 601171, 601026, 601122, 601208, 601103, 601089, 601108, 601183, 600933 600971, 600919, 600910, 600904, 600935, 601005, 600977, 600936, 600993, 600965, 601011, 600957, 600918, 601015, 600992, 600905, 600914, 600947, 601016, 601013, 600981, 600903, 600949, 601012, 600983, 600970, 600912, 601004, 600915, 600907, 600987, 600939, 600996, 601007, 600899, 601006, 600938, 600952, 600988, 601020, 600928, 600937, 600972, 955056, 965059, 954609, 949471, 954490, 501442, and 499300.