Stryker Advanced Cement Mixer (ACM) Bowl and Base without 180 Gram Cement Cartridge for use with Stryker High...

Stryker Corp. / initiated 2008-01-10 / Terminated / root cause: manufacturing process

Stryker Instruments Div. of Stryker Corporation began a recall of Stryker Advanced Cement Mixer (ACM) Bowl and Base without 180 Gram Cement Cartridge for use with Stryker High Vacuum Cement Injection System, sterile; REF... on . The reason FDA records is "Sterility may be compromised, as the packaging may have channels in the packaging seal". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1106-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
Stryker Advanced Cement Mixer (ACM) Bowl and Base without 180 Gram Cement Cartridge for use with Stryker High Vacuum Cement Injection System, sterile; REF 306-554, Stryker Instruments, Kalamazoo, MI 49001.
Product code
JDZ Mixer, Cement, For Clinical Use
Reason for recall
Sterility may be compromised, as the packaging may have channels in the packaging seal.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Stryker notified consignees via Medical Device Recall Notification letter dated 1/10/08 to quarantine the product and to return it.
Quantity in commerce
25 boxes of 6 each
Distribution
Nationwide Distribution - Illinois, Michigan, New York, Ohio and Wyoming.
Product codes and lots
Lot Numbers: 05167CG2, 05192012, 05223012 and 06335012.