Z-1186-2008

Biomet brand Modular Microplasty Cup Impactor, 3/8" - 24 Thread Insert; Model 31-400603. Instrument for hip...

Zimmer GmbH / initiated 2008-01-16 / Terminated / root cause: design

Biomet, Inc. began a recall of Biomet brand Modular Microplasty Cup Impactor, 3/8" - 24 Thread Insert; Model 31-400603. Instrument for hip prosthesis acetabular cup component on . The reason FDA records is "The weld at the lock location may fracture during impaction". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1186-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Biomet brand Modular Microplasty Cup Impactor, 3/8" - 24 Thread Insert; Model 31-400603. Instrument for hip prosthesis acetabular cup component.
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
The weld at the lock location may fracture during impaction.
Root cause tag
design
FDA root cause
Device Design
Action
Biomet notified distributors and medical facilities by letter dated 1/16/08. Distributors with distributor-owned instruments at medical facilities were notified by email on 1/17/08, and provided with a letter to present to their associated medical facilities describing the recall.
Quantity in commerce
59
Distribution
Worldwide Distribution: USA, Germany, Mexico, Netherlands, Sweden and United Kingdom.
Product codes and lots
Lot 095207.