Z-1050-2008

Physio-Control LIFEPAK 20 defibrillator/monitor, AC powered external defibrillator with battery backup. Part...

Physio Control, Inc. / initiated 2008-01-08 / Terminated / root cause: manufacturing process

Physio Control, Inc. began a recall of Physio-Control LIFEPAK 20 defibrillator/monitor, AC powered external defibrillator with battery backup. Part Numbers: 3202487-073, 3202487-079, 3202488-036... on . The reason FDA records is "Inoperable Device - Potential for delay in therapy or prevention of defibrillation therapy due to corrosion of the Printed Circuit Board Assemblies (PCBA)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1050-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Physio Control, Inc.
Product
Physio-Control LIFEPAK 20 defibrillator/monitor, AC powered external defibrillator with battery backup. Part Numbers: 3202487-073, 3202487-079, 3202488-036, Physio Control, Inc, Redmond, WA 98052
Product code
MKJ Automated External Defibrillators (Non-Wearable)
Reason for recall
Inoperable Device - Potential for delay in therapy or prevention of defibrillation therapy due to corrosion of the Printed Circuit Board Assemblies (PCBA).
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm contacted Customers by phone and by Urgent-Medical Device Recall letters sent via return receipt requested to their consignees on 1/8/08. The letter advised consignees that the Printed Circuit Board Assemblies (PCBA) might not meet quality requirements. This manufacturing deviation could lead to eventual PCBA corrosion that could render the device inoperative and not ready for use in an emergency under battery power. The firm recommended keeping the unit in service until it could be replaced. Representatives will visit each consignee and replace units.
Quantity in commerce
103 total devices - 84 domestic, 19 international
Distribution
Worldwide Distribution --- USA including states of OK, SC, TX, and WA, and country of Germany.
Product codes and lots
Serial Numbers (domestic units): 36133753, 36133760, 36139018, 36139023, 36139030, 36137448, 36138998, 36141356, 36141357, 36141358, 36141360, 36141361, 36141362, 36141363, 36141365, 36141366, 36141367, 36141370, 36141373, 36141378, 36141380, 36141381, 36141382, 36141383, 36141384, 36141393, 36141394, 36141395, 36141396, 36141399, 36141400, 36142717, 36142718, 36142730, 36142731, 36141386, 36141398, 36133739, 36133740, 36133742, 36133743, 36133744, 36133745, 36133746, 36133747, 36133748, 36133749, 36133750, 36133751, 36133752, 36133754, 36133755, 36133756, 36133757, 36133758, 36133759, 36137218, 36137219, 36137220, 36137222, 36137224, 36137225, 36137226, 36137227, 36137230, 36137231, 36137352, 36137353, 36137355, 36137356, 36137357, 36137359, 36137360, 36137361, 36137362, 36137363, 36137364, 36137365, 36137366, 36137367, 36137368, 36137369, 36139002, 36139010  Serial Numbers (international units): 36137354, 36137358, 36137370, 36137371, 36137372, 36138999, 36139000, 36139005, 36139006, 36139007, 36139009, 36139011, 36139012, 36139013, 36139014, 36139021, 36139024, 36139025, 36139031