Physio-Control LIFEPAK 20 defibrillator/monitor, AC powered external defibrillator with battery backup. Part...
Physio Control, Inc. began a recall of Physio-Control LIFEPAK 20 defibrillator/monitor, AC powered external defibrillator with battery backup. Part Numbers: 3202487-073, 3202487-079, 3202488-036... on . The reason FDA records is "Inoperable Device - Potential for delay in therapy or prevention of defibrillation therapy due to corrosion of the Printed Circuit Board Assemblies (PCBA)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1050-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Physio Control, Inc.
- Product
- Physio-Control LIFEPAK 20 defibrillator/monitor, AC powered external defibrillator with battery backup. Part Numbers: 3202487-073, 3202487-079, 3202488-036, Physio Control, Inc, Redmond, WA 98052
- Product code
- MKJ Automated External Defibrillators (Non-Wearable)
- Reason for recall
- Inoperable Device - Potential for delay in therapy or prevention of defibrillation therapy due to corrosion of the Printed Circuit Board Assemblies (PCBA).
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The firm contacted Customers by phone and by Urgent-Medical Device Recall letters sent via return receipt requested to their consignees on 1/8/08. The letter advised consignees that the Printed Circuit Board Assemblies (PCBA) might not meet quality requirements. This manufacturing deviation could lead to eventual PCBA corrosion that could render the device inoperative and not ready for use in an emergency under battery power. The firm recommended keeping the unit in service until it could be replaced. Representatives will visit each consignee and replace units.
- Quantity in commerce
- 103 total devices - 84 domestic, 19 international
- Distribution
- Worldwide Distribution --- USA including states of OK, SC, TX, and WA, and country of Germany.
- Product codes and lots
- Serial Numbers (domestic units): 36133753, 36133760, 36139018, 36139023, 36139030, 36137448, 36138998, 36141356, 36141357, 36141358, 36141360, 36141361, 36141362, 36141363, 36141365, 36141366, 36141367, 36141370, 36141373, 36141378, 36141380, 36141381, 36141382, 36141383, 36141384, 36141393, 36141394, 36141395, 36141396, 36141399, 36141400, 36142717, 36142718, 36142730, 36142731, 36141386, 36141398, 36133739, 36133740, 36133742, 36133743, 36133744, 36133745, 36133746, 36133747, 36133748, 36133749, 36133750, 36133751, 36133752, 36133754, 36133755, 36133756, 36133757, 36133758, 36133759, 36137218, 36137219, 36137220, 36137222, 36137224, 36137225, 36137226, 36137227, 36137230, 36137231, 36137352, 36137353, 36137355, 36137356, 36137357, 36137359, 36137360, 36137361, 36137362, 36137363, 36137364, 36137365, 36137366, 36137367, 36137368, 36137369, 36139002, 36139010 Serial Numbers (international units): 36137354, 36137358, 36137370, 36137371, 36137372, 36138999, 36139000, 36139005, 36139006, 36139007, 36139009, 36139011, 36139012, 36139013, 36139014, 36139021, 36139024, 36139025, 36139031