Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1031-2008

Stryker 5.0mm Round Fluted Bur, Extra Long, for use with TPS Universal Drill Extra Long "R" attachment...

Stryker Corp. / initiated 2008-01-10 / Terminated / root cause: packaging

Stryker Instruments Div. of Stryker Corporation began a recall of Stryker 5.0mm Round Fluted Bur, Extra Long, for use with TPS Universal Drill Extra Long "R" attachment, sterile, Stryker Instruments, Kalamazoo, MI; REF... on . The reason FDA records is "Lack of assurance of sterility". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1031-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
Stryker 5.0mm Round Fluted Bur, Extra Long, for use with TPS Universal Drill Extra Long "R" attachment, sterile, Stryker Instruments, Kalamazoo, MI; REF 5260-10-50.
Product code
HBE Drills, Burrs, Trephines & Accessories (Simple, Powered)
Reason for recall
Lack of assurance of sterility.
Root cause tag
packaging
FDA root cause
Package design/selection
Action
Stryker sent a Medical Device Recall Notification letter dated 1/10/08 to consignees. The letter instructed them to check their facility for product, cease using it, and to return the product using a mailing label provided. International subsidiaries were notified via email.
Quantity in commerce
10,588 of all products
Distribution
Worldwide Distribution - USA, Australia, Brazil, Canada, England, France, Germany, Hong Kong, India, Italy, Japan, Mexico, the Netherlands, New Zealand, Portugal, Saudi Arabia, Poland, Romania, South Africa, Spain, Sweden, Switzerland and United Arab Emirates.
Product codes and lots
All lots.