Z-1037-2008

Straumann NC Closure Screw, Titanium, Precision attachment, Article Number: 024.2105.04, Straumann, Andover...

Straumann USA, LLC / initiated 2008-01-10 / Terminated / root cause: labeling and UDI

Straumann Usa, Llc began a recall of Straumann NC Closure Screw, Titanium, Precision attachment, Article Number: 024.2105.04, Straumann, Andover, MA 01810 on . The reason FDA records is "Mislabeled package: Narrow Connection (NC) Closure Screws contain closure screws for the regular connection (RC) bone level implant. The RC closures screw is larger (3.5mm diameter) and therefore, will not fit into the...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1037-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Straumann USA, LLC
Product
Straumann NC Closure Screw, Titanium, Precision attachment, Article Number: 024.2105.04, Straumann, Andover, MA 01810
Product code
EGG Attachment, Precision, All
Reason for recall
Mislabeled package: Narrow Connection (NC) Closure Screws contain closure screws for the regular connection (RC) bone level implant. The RC closures screw is larger (3.5mm diameter) and therefore, will not fit into the NC implant (3.0mm diameter) narrow connection.
Root cause tag
labeling and UDI
FDA root cause
Mixed-up of materials/components
Action
Consignees were notified by telephone on 1/10/08 followed by an Urgent Device Recall letter on January 10, 2008. The letter instructs users to return any affected product to the firm along with a recall acknowledgement form. For additional information, contact 978-747-2514.
Quantity in commerce
26 units
Distribution
Nationwide Distribution including states of CA, DE, FL, IL, ME, NY, OH, OR, and TX.
Product codes and lots
Lot Number: G8677