Z-1530-2008
BD FACSDiva Software v6.0 and 6.1 with the HP XW4400 Computer System Model/Catalog number 642685 Software is...
Becton, Dickinson and Company, BD Biosciences / initiated 2007-12-21 / Terminated / root cause: software
BD Biosciences began a recall of BD FACSDiva Software v6.0 and 6.1 with the HP XW4400 Computer System Model/Catalog number 642685 Software is used to simplify acquisition of flow cytometry data on . The reason FDA records is "Values might not update in statistics views after certain functions are performed in a worksheet". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1530-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton, Dickinson and Company, BD Biosciences
- Product
- BD FACSDiva Software v6.0 and 6.1 with the HP XW4400 Computer System Model/Catalog number 642685 Software is used to simplify acquisition of flow cytometry data.
- Product code
- GKZ Counter, Differential Cell
- Reason for recall
- Values might not update in statistics views after certain functions are performed in a worksheet.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm issued an Important Product Information notification letter on 12/21/2007. The letter instructs users to reload data by closing and reopening an experiment before reporting results in order to determine if their statistics have been impacted. They are asked to fill out and return a form pertaining to this problem. BD Biosciences will release a version of BD FACSDiva software to correct this problem. The letter also stated that BD Biosciences rep will contact the users when the new software is available. User are asked to contact BD Customer Support Center at 1-877-232-8995 if they have any additional questions.
- Quantity in commerce
- 74 units
- Distribution
- Nationwide Distribution.
- Product codes and lots
- Material numbers 338960, 338962, and 339473