Z-1244-2008
WEB1000 Versions 3.1, 4.1 and 5.1, Picture Archiving and Communications systems (PACs), used in the...
AGFA Healthcare Corp. / initiated 2007-05-14 / Terminated / root cause: software
AGFA Corp. began a recall of WEB1000 Versions 3.1, 4.1 and 5.1, Picture Archiving and Communications systems (PACs), used in the acceptance, transfer, display, storage and digital... on . The reason FDA records is "Java version 1.6x (JAVA 6) is not compatible with AGFA IMPAX Administration/Service Tools. Due to a Software Mismatch data can be lost". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1244-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- AGFA Healthcare Corp.
- Product
- WEB1000 Versions 3.1, 4.1 and 5.1, Picture Archiving and Communications systems (PACs), used in the acceptance, transfer, display, storage and digital processing of medical images.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Java version 1.6x (JAVA 6) is not compatible with AGFA IMPAX Administration/Service Tools. Due to a Software Mismatch data can be lost.
- Root cause tag
- software
- FDA root cause
- Software change control
- Action
- Agfa Healthcare contacted all consignees on 5/14/2007 and by registered mail on 06/07/2007. The registered Urgent Field Safety Notice letters requested the accounts take the following steps immediately: 1. Turn off automatic JAVA updates to avoid this update being applied in the background. 2. Do not manually update JAVA on their station-even for other applications. 3. If JAVA 1.6 has already been applied, do not use Service/Admin Tools. They were instructed to contact their IT department or Agfa support for assistance in backing out of this version to apply the required version for the Agfa software. A Mandatory Service Bulletin will inform and assist affected account in removing JAVA 1.6 if it has been applied and turning off the automatic updates via the JAVA Plug-in Control Panel.
- Quantity in commerce
- 946 units (total)
- Distribution
- U.S. Nationwide and Canada.
- Product codes and lots
- WEB1000 version 3.1, Catalog Number: LD5GQ001; WEB1000 version 4.1, Catalog Number: LZS5R, LZS9Z, LZTA4 and WEB1000 version 5.1, Catalog Number: L7DOS, L7DLPU.