Z-1447-2008

Given Diagnostic System with PillCam SB Capsule - Gastrointestinal capsule imaging system. The device is an...

Given Imaging Ltd. / initiated 2007-12-13 / Terminated / root cause: manufacturing process

Given Imaging Inc began a recall of Given Diagnostic System with PillCam SB Capsule - Gastrointestinal capsule imaging system. The device is an ingestible telemetric gastrointestinal capsule... on . The reason FDA records is "The cards included in the product's education kit (proposing a clinical algorithm for capsule endoscopy in patients with suspected Crohn's disease), were printed incorrectly". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1447-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Given Imaging Ltd.
Product
Given Diagnostic System with PillCam SB Capsule - Gastrointestinal capsule imaging system. The device is an ingestible telemetric gastrointestinal capsule imaging system is used for visualization of the small bowel mucosa as an adjunctive tool in the detection of abnormalities of the small bowel. The device captures images of the small bowel with a wireless camera contained in a capsule. This device includes an ingestible capsule (containing a light source, camera, transmitter, and battery), an antenna array, a receiving/recording unit, a data storage device, computer software to process the images, and accessories.
Product code
NEZ System, Imaging, Gastrointestinal, Wireless, Capsule
Reason for recall
The cards included in the product's education kit (proposing a clinical algorithm for capsule endoscopy in patients with suspected Crohn's disease), were printed incorrectly.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Consignees were notified by letter on 12/13/2007. The letter contains a diagram illustrating the error on the incorrect card and a diagram illustrating the corrected card. Consignees were instructed to discard the incorrect card if they still have it and to complete a response form and return it via fax to Given Imaging.
Quantity in commerce
1392 units
Distribution
Worldwide Distribution --- US including Puerto Rico and country of Canada
Product codes and lots
GMB-0253-01-GLR Card-Proposed Diagnostic Algorithm for Suspected Crohn's Disease.