Z-1140-2008

B-K Medical Intraoperative Transducer Type 8814, diagnostic ultrasound transducer, B-K Medical Systems, Inc...

B & K Medical Systems, Inc. / initiated 2008-01-24 / Terminated / root cause: manufacturing process

B & K Medical Systems, Inc. began a recall of B-K Medical Intraoperative Transducer Type 8814, diagnostic ultrasound transducer, B-K Medical Systems, Inc., Wilmington, MA 01887 on . The reason FDA records is "Crack: Incomplete glue joint in the device housing may affect electrical safety or sterilization". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1140-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
B & K Medical Systems, Inc.
Product
B-K Medical Intraoperative Transducer Type 8814, diagnostic ultrasound transducer, B-K Medical Systems, Inc., Wilmington, MA 01887
Product code
IYN System, Imaging, Pulsed Doppler, Ultrasonic
Reason for recall
Crack: Incomplete glue joint in the device housing may affect electrical safety or sterilization.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
B-K Medical issued letter notification on 1/23/08 to accounts via Fed Ex, advising them not to use the unit until a representative can inspect and exchange, if unit is found defective. Instructions are also provided to continue use of the device with a sterile cover until site visit.
Quantity in commerce
7
Distribution
USA Distribution: MA, MI, PA, TX, and OR.
Product codes and lots
Serial Numbers: 1864742, 1869148, 1869619, 1870824, 1870825, 1872061, and 1875317.