Z-1140-2008
B-K Medical Intraoperative Transducer Type 8814, diagnostic ultrasound transducer, B-K Medical Systems, Inc...
B & K Medical Systems, Inc. / initiated 2008-01-24 / Terminated / root cause: manufacturing process
B & K Medical Systems, Inc. began a recall of B-K Medical Intraoperative Transducer Type 8814, diagnostic ultrasound transducer, B-K Medical Systems, Inc., Wilmington, MA 01887 on . The reason FDA records is "Crack: Incomplete glue joint in the device housing may affect electrical safety or sterilization". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1140-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- B & K Medical Systems, Inc.
- Product
- B-K Medical Intraoperative Transducer Type 8814, diagnostic ultrasound transducer, B-K Medical Systems, Inc., Wilmington, MA 01887
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- Crack: Incomplete glue joint in the device housing may affect electrical safety or sterilization.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- B-K Medical issued letter notification on 1/23/08 to accounts via Fed Ex, advising them not to use the unit until a representative can inspect and exchange, if unit is found defective. Instructions are also provided to continue use of the device with a sterile cover until site visit.
- Quantity in commerce
- 7
- Distribution
- USA Distribution: MA, MI, PA, TX, and OR.
- Product codes and lots
- Serial Numbers: 1864742, 1869148, 1869619, 1870824, 1870825, 1872061, and 1875317.