Z-0983-2008
VIASYS AVEA Ventilator, Catalog numbers: 17210-03 (Italian Comprehensive Version)
ZOLL Palm Springs / initiated 2007-12-17 / Terminated / root cause: design
Viasys Respiratory Care, Inc.dba Bird Products began a recall of VIASYS AVEA Ventilator, Catalog numbers: 17210-03 (Italian Comprehensive Version) on . The reason FDA records is "Alarm Failure -- There is a potential of alarms not being activated subsequent to specific electronic faults that may cause INOP condition". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0983-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- ZOLL Palm Springs
- Product
- VIASYS AVEA Ventilator, Catalog numbers: 17210-03 (Italian Comprehensive Version)
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- Alarm Failure -- There is a potential of alarms not being activated subsequent to specific electronic faults that may cause INOP condition.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- On December 18, 2007, a product notification letter was mailed to all customers in the United States via overnight carrier (registed mail) and a product notification letter was mailed to all international customers via Federal Express. The recall letter informed the customers of the problem and also inform them that a Cardinal Health (Viasys) service representative will be contacting them immediately to arrange & service their ventilator. For futher information please contack Andrea von Muller at 760-778-7323.
- Quantity in commerce
- 1
- Distribution
- Worldwide Distribution --- USA including states of AL, CA, MA, NM, NV, and NY and country of Italy.
- Product codes and lots
- Catalog numbers: 17210-03; Serial numbers: AGV01472